Syringe Pump

Infusion Syringe Pump Systems







Product Features

  • Continuous Infusion and TCI (Target Controlled Infusion) Modes
  • Preset ('hands-free') and Manual ('hands-on') Bolus
  • High Bolus Infusion Rate
  • Volume (ml/hr) and Mass Units Infusion Programming
  • Accepts and automatically senses the size of syringes from the major manufactures
  • Occlusion Alarm
  • Volume and Mass and Totalizers
  • Pump History
  • Communications Interface
  • AC Powered with backup battery operation
  • TCI Features
    • Incorporates the AstraZeneca DiprifusorTM TCI Module
    • Compatible with AstraZeneca's Diprivan PFS
    • TCI Target Titrate


PRODUCT SPECIFICATIONS


Infusion Modes

  • TCI and Continuous

Programming Modes

  • TCI target (blood) concentration, volume and mass units

Continuous Infusion Rate Range

  • 0.1 - 400.0 ml/hr in 0.1 ml increments

Bolus Infusion Rate Range

  • 0.1 - 1200.0 ml/hr in 0.1 ml increments

Mass Unit Programming Units

  • mg/hr, mg/kg/hr, mg/kg/min, mcg/hr, mcg/kg/hr, mcg/kg/min

Dipifusor™ Patient Details

  • Age: 16-100
  • Weight: 30-150kg
  • Target: 0.1 - 15 mcg/ml

Syringe Types

  • BD Plastipak 5, 10, 20, 30, 50/60 ml
  • Braun Omnifix 5, 10, 20, 30, 50/60 ml
  • Monoject 5, 10, 20, 30/35, 50/60 ml
  • Terumo 5, 10, 20, 30, 50/60 ml
  • Fresenius 50 ml
  • IMS Pumpjet 30 ml (prefilled)
  • Zeneca PFS 50 ml (1% and 2%)
  • Nipro 5, 10, 20, 30, 50/60 ml
  • JMS 5, 10, 20, 30, 50/60 ml

Drive Accuracy

  • The delivery volume is accurate to within ±2% over a complete syringe length (excluding the startup time)

TOTALIZERS

TCI Mode

  • 0 – 999.9 ml (in 0.1 ml increments)
  • 0 – 99999.9 mg (in 0.1 mg increments)

Normal Mode

  • 0 – 999.9 ml (in 0.1 ml increments)
  • 0 – 99999.999 mg (in 0.001 mg increments)

History

  • Log of up to 1,500 events

Communications Interface

  • RS232 interface, communications protocol available on request

AC Power Supply

  • 100-240V at 50/60Hz, 40VA

Battery Type

  • Sealed Lead Acid (Cyclon)

Battery Life

  • 10 hours (fully charged and under normal operating conditions)

Electrical Safety Classification

  • Class II, Type CF

Fluid Ingress

  • IPX1 protection against drops of water falling vertically

Dimension

  • 325 x 195 x 115 mm (with pole clamp fitted and plunger clamp closed)

Weight

  • Not exceeding 3.5 kg including batteries and pole clamp

Operating Conditions

  • +5 to +40°C, 30 to 75% RH, 700 to 1060 hPa

Design standards

  • IEC 601-1, IEC 601-1-2

© Copyright 2018. Medicare Electronics, All Rights Reserved.
Designed and Development Shivam IT Solution.

Office: 183. Pkt-3, Sector-24, Rohini, New Delhi - 110085,
Mob: 09210333575, 09873856002